Specifications of Bruker Avance

800 and 600 MHz NMR Spectrometers

 

For the specifications of the TCI CryoProbe, please click here.

Magnet

 

 

 

 

 

 

 

Description

Specification for Avance 800

Specification for Avance 600

 

 

 

 

 

Drift of field

2.33 Hz/h

5 Hz/h

 

 

 

 

 

Lineshape @ 0.55%/0.11%,without lock (ns = 8,over 8 h)

Hump = 13.3  Hz  / 24.7 Hz

 

 

 

 

 

 

Liquid Helium

260 liters holding for  >58 days

60 liters holding for >50 days

 

 

 

 

 

Temperature of liquid helium

1.92-1.93  K

4.2 K

 

 

 

 

 

Liquid Nitrogen

230 liters holding for > 12 days

160 liters holding for >13 days

 

 

 

 

 

 

 

 

 

 

 

 

 

TXI probe

 

 

 

 

 

 

 

Experiment description

Specification for Avance 800

Specification for Avance 600

 

 

 

 

 

Sensitivity of 1H

1879:1 or 1746:1 for CH2 group

908:1

 

 

 

 

 

Sensitivity of 13C

 

 

 

 

 

 

 

Sensitivity of 15N

 

 

 

 

 

 

 

Lineshape without rotation @ 0.55%/0.11%

Hump = 6.9 Hz / 13.5 Hz

6.7 Hz / 11.7 Hz

 

 

 

 

 

Lineshape with rotation @ 0.55%/0.11%

Hump = 5.0 Hz / 6.9 Hz, spinning sideband height = 0.4%

4.5 Hz / 7.6 Hz, Spinning sideband height = 0.1%

 

 

 

 

 

Resolution of 1H

0.46 Hz

0.42 Hz

 

 

 

 

 

Water suppression

Hump = 40.8 / 80.5 Hz @ pl9 = 50 dB, Splitting = 24%, Sino = 467

92.8  / 160.3 Hz @pl9 = 54dB,Splitting = 22%, Sino = 173

 

 

 

 

 

1H 90 degree pulse width of high power

9us @ -3dB, 19 Watts

8us @ -1.2 dB

 

 

 

 

 

13C high power decoupling pulse

10.2us @ -6dB, 480 Watts

12.3 us @ -4 dB

 

 

 

 

 

13C low power decoupling pulse

75 us @ 13 dB

65 us @ 12 dB

 

 

 

 

 

15N high power decoupling pulse

30 us @ -6 dB, 480 Watts

38 us @ -2.8 dB

 

 

 

 

 

15N low power decoupling pulse

190 us @ 11.2 dB

200 us @ 12 dB

 

 

 

 

 

 

 

 

 

 

 

 

 

QXI probe

 

 

 

 

 

 

 

Experiment description

Specification for Avance 800

Specification for Avance 600

 

 

 

 

 

Sensitivity of 1H

1157:1 or 1336:1 for CH2 group

 943:1 or 1040:1 for CH2 group

 

 

 

 

 

Sensitivity of 13C

90

 125 for 10% EB in CDCl3

 

 

 

 

 

Sensitivity of 15N

14

 

 

 

 

 

 

Sensitivity of 31P

45

 92 for 0.0485 M TPP in acetone-d6

 

 

 

 

 

Lineshape without rotation @ 0.55%/0.11%

6.2 Hz / 10.1 Hz

  5.0 Hz/11.9 Hz

 

 

 

 

 

Lineshape with rotation @ 0.55%/0.11%

3.7 Hz / 7.0 Hz; Spinning sideband height = 0.4%

 

 

 

 

 

 

Resolution of 1H

0.48 Hz

 0.72 Hz

 

 

 

 

 

Water suppression

Hump = 45.8 Hz / 88.7 Hz @ pl9 = 55 dB, Resolution = 23%, Sino = 305

Hump = 29.1 Hz/42.6 Hz @ pl9=58 dB, Resolution=16%, Sino=202  

 

 

 

 

 

1H high power decoupling pulse

9.5 us @ -4  dB, 20 Watts

 

 

 

 

 

 

1H low power decoupling pulse

35 us @ 15 dB

 

 

 

 

 

 

13C high power decoupling pulse

15.2 us @ 0 dB, ? Watts

 

 

 

 

 

 

13C low power decoupling pulse

73us @ 13 dB

 

 

 

 

 

 

15N high power decoupling pulse

25us @ -5dB, 480 Watts

 

 

 

 

 

 

15N low power decoupling pulse

190 us @ 13.5 dB

 

 

 

 

 

 

31P high power decoupling pulse

25 us @ 2.3 dB, 120 Watts

 

 

 

 

 

 

 

                                               The Sensitivity comparisons of the first FIDs of

                                         2D and 3D experiments on Bruker Avance 800 and 600

  

 

 

 

 

 

 

 

 

 

 

                                                              Back to Top